Core Functions and Technical Mechanisms of Sensors in Thermal Shock Test Chambers ——With a Discussion on Their Decisive Impact on Test Accuracy and Reliability

Source:LINPIN Time:2025-09-23 Category:Industry News

Introduction
In the field of environmental reliability testing, thermal shock test chambers are widely used to evaluate the adaptability and lifetime degradation of electronic components, materials, and complete products under rapid and extreme temperature transients. Whether the chamber can faithfully reproduce the temperature-profile specified by the standard depends not only on the design margin of the refrigeration/heating system, but also on the cabinet structure, air-duct layout and thermal insulation scheme. All of these hardware links, however, must ultimately be regulated through the closed-loop chain of “sensor–controller–actuator”. As the first element of the information chain, the sensor’s performance directly determines the equipment’s indication error, overshoot amplitude, recovery time and long-term stability. Written in a formal technical style, this paper systematically expounds the operating principles, selection criteria, installation essentials and error-control strategies of sensors commonly used in thermal shock test chambers (thermocouples, platinum resistance thermometers, infrared radiation thermometers, digital bus sensors, etc.). Combined with the latest metrology standards and industrial cases, engineering recommendations for improving test accuracy are proposed for equipment developers, metrologists and test engineers.


Functional Positioning of Sensors in Test Chambers
Original data acquisition unit for temperature values
Sensors convert the invisible temperature field inside the working space into quantifiable and transmittable electrical parameters, forming the source for all subsequent control, recording, alarm and traceability operations.
Feedback reference for control algorithms
PID, fuzzy-PID, Model Predictive Control (MPC) and other algorithms all take the real-time sampled value of the sensor as the feedback variable. If the reference is distorted, even the most advanced algorithm cannot eliminate the systematic error.
Threshold monitor for the safety chain
Failures such as over-temperature, under-temperature, excessive rate of change, or sensor wire-break first manifest themselves as abrupt changes in the sensor signal. The sensor therefore assumes the dual safety duty of protecting both the equipment and the specimen.
Link for metrological traceability
According to ISO/IEC 17025 and GB/T 19022, the chamber must periodically undergo calibration of temperature deviation, fluctuation and uniformity. As the bridge between the unit under test and the standard, the integrity of the sensor’s traceability chain determines the international mutual recognition of the calibration result.
Mainstream Sensor Types and Measurement Principles
3.1 Thermocouple
a. Seebeck effect
Two dissimilar conductors A and B form a closed circuit. When their junctions are kept at different temperatures T? and T?, a thermo-electromotive force E_AB is generated:
E_AB = α(T? – T?) + 0.5β(T?2 – T?2)
where α and β are material-specific coefficients.
b. Common types and ranges
Type K (–200 °C to +1 250 °C), Type N (–200 °C to +1 300 °C) and Type T (–200 °C to +350 °C) are mainstream choices for thermal-shock chambers. Type K, featuring good oxidation resistance, moderate linearity and low cost, accounts for more than 70 % of applications.
c. Cold-junction compensation
According to the law of intermediate temperatures, any fluctuation in cold-junction temperature produces a systematic offset. Modern acquisition modules commonly adopt a PT1000 cold-junction sensor plus a digital temperature chip, keeping the compensation error within ±0.1 °C.
3.2 Platinum Resistance Thermometer (RTD, Pt100/Pt1000)
Based on the positive temperature coefficient of platinum, IEC 60751 stipulates:
R(t) = R?(1 + At + Bt2 + Ct3)
where R? is the resistance at 0 °C and A, B, C are constants. Pt100 offers excellent reproducibility and long-term stability (annual drift ≤ 0.05 °C) in the –70 °C to +200 °C range, making it suitable for the “reference channel” of high-precision chambers.
3.3 Infrared Radiation Thermometer (IRT)
For non-contact, fast-moving specimen surfaces (e.g. semiconductor wafers, relay housings), thermopile detectors in the 8–14 μm band are used, giving a t?? response time ≤ 50 ms and avoiding the thermal-inertia error caused by thermocouple mass loading. Main error sources are emissivity setting, field-of-view contamination and ambient reflection, requiring black-body calibration and algorithmic correction.
3.4 Digital Bus Sensors (I2C, RS-485, CAN)
New-generation sensors integrate ADC, calibration coefficients and linearisation algorithms inside a TO-92 or stainless-steel probe, outputting a digital temperature value directly. This eliminates noise from long-line transmission and thermocouple inhomogeneity, supports TEDS (Transducer Electronic Data Sheet) automatic identification, and facilitates the construction of smart laboratories.
Influence Mechanisms of Sensors on Test Accuracy
4.1 Time constant τ and temperature overshoot
τ = ρc?V/(hA), where ρ is density, c? specific heat, V probe volume, h convection coefficient and A surface area. A larger τ means slower response, causing the controller to “think” the target has not been reached and to continue heating/cooling, thus producing overshoot. Since the transition cycle in a thermal-shock chamber is usually ≤ 5 min, τ ≤ 10 s is required. In practice, exposed-junction Type K wires of φ0.5 mm or thin-film Pt100 are used, and τ is further reduced by decreasing sheath diameter and increasing flow velocity.
4.2 Heat-conduction error
If the probe touches the wall or is mounted on a metal bracket, an additional heat bridge is introduced, making the indicated temperature deviate from the air temperature. Low-conductivity PEEK or PTFE brackets should be used, and the sensing point must be kept ≥ 20 mm from any wall.
4.3 Self-heating error
For RTDs, the excitation current I produces Joule heat Q = I2R. If heat dissipation is poor, the reading will be high. IEC 60751 recommends a self-heating coefficient ≤ 0.2 °C mW?1 for Pt100; a 1 mA constant-current source can meet the ±0.05 °C accuracy requirement.
4.4 Electromagnetic compatibility (EMC)
High-power compressors and solid-state relays generate dv/dt and di/dt during switching, which can couple into the millivolt-level thermocouple signal through inter-wire capacitance and mutual inductance, causing spike interference. Countermeasures include shielded twisted-pair cable, differential amplification, hardware RC plus digital FIR filtering, and floating-ground techniques.
4.5 Uniformity and loading effects
GB/T 2423.22 requires temperature uniformity ≤ 2 °C (–65 °C to +150 °C). Sensors must be placed at nine points: three vertical levels, each with four corners plus the centre. Poor sensor consistency will amplify or mask chamber airflow defects, leading to misjudgement.
Engineering Practice for Sensor Selection and Positioning
Accuracy class
The expanded uncertainty (k = 2) after calibration should be ≤ 1/3 of the test-allowable error. For example, if the specification permits ±2 °C deviation, sensor calibration uncertainty must be ≤ 0.7 °C.
Redundancy strategy
For critical tests (aerospace-grade parts), a “two-out-of-three” voting logic is adopted: three identical thermocouples are installed side-by-side; the controller compares them in real time and shuts down immediately if any one deviates beyond the limit.
Quick-lock structure
M12 stainless-steel aviation connectors plus extension compensating cables facilitate periodic calibration and avoid Seebeck-coefficient drift caused by repeated terminal tightening.
Anti-condensation and pressure-proof
During low-temperature shock, the dew-point difference often exceeds 30 °C. Sensor sheaths must be sealed to IP67, with molecular-sieve desiccant inside the junction box. Liquid-nitrogen-cooled chambers must withstand 0.4 MPa internal pressure, requiring sheath wall thickness ≥ 0.3 mm.
Latest Standards and Metrological Trends
GB/T 5170.10-2021 “Methods for Verification of Environmental Testing Equipment—Thermal Shock Test Chambers” specifies:
a) Calibration points shall cover minimum, maximum and mid-range;
b) Sensor sampling frequency ≥ 1 Hz, recording interval ≤ 30 s;
c) Cold-junction compensation method and uncertainty budget shall be documented.
IEC 60584-1:2021 tightens the reference function for thermocouples; the maximum deviation of Type K in the –200 °C to 0 °C range is reduced from ±2.2 °C to ±1.5 °C, and the updated table became mandatory on 1 Jan 2023.
JJG (Aerospace) ×××-2025 (draft) introduces the concept of “dynamic temperature deviation”: under a rate of 5 °C min?1, the 30 s sliding average deviation shall be ≤ ±1 °C, imposing stricter requirements on sensor response and digital filtering.
Failure Cases and Improvement Measures
Case 1: A third-party laboratory experienced premature specimen failure during –55 °C ? +125 °C cycling. Root cause: micro-cracks in the Type K sheath allowed moisture ingress, causing a +1.8 °C drift in thermal EMF. Improvement: replace with 316 L MgO-insulated mineral-insulated cable and perform 10 MPa helium leak testing.
Case 2: During communication-power-module tests, the controller displayed ±6 °C spikes at switching instants. Investigation revealed beat-frequency interference between the compressor inverter carrier and the acquisition card sampling rate. Improvement: add a second-order RC passive filter (f_c = 10 Hz) at the thermocouple input and shift sampling frequency to a non-integer 9.7 Hz, reducing spikes to ±0.3 °C.
Conclusion
Sensors in thermal shock test chambers are far from simple “temperature probes”; they are precision front ends that integrate materials science, thermodynamics, electronic measurement and data science. As third-generation semiconductors, automotive-grade chips, low-orbit satellites and other high-end applications demand temperature-profile fidelity of ≤ ±0.5 °C or even ±0.3 °C, sensor selection, positioning, calibration and maintenance must be embedded in the core of the chamber’s life-cycle quality management. In the future, emerging technologies such as MEMS micro-thermocouple arrays with AI self-calibration, quantum-cascade-laser absorption spectroscopy (QCLAS) for non-contact thermometry, and digital-twin real-time mapping will further propel thermal-shock testing from “empirical replication” to “digital precision”, providing rock-solid metrological support for the reliability verification of high-end equipment.

News Recommendation
Safety for any test chamber is very important, now the equipment are set up to protect the measures, in addition to these self-protection warning, in the usual use of equipment in the process also need to pay attention to the following points.
Regularly, staff should master some troubleshooting and handling methods for high and low temperature alternating test chambers. If the equipment exhibits unstable low temperatures, what causes this phenomenon, and how should we address these faults?
Previously, we discussed the changes that occur in objects during high-temperature testing. But what happens when materials are subjected to low-temperature testing in a low-temperature test chamber?
A low-temperature test chamber can accommodate multiple test specimens, but are there specific requirements regarding the location and quantity of test points? Today, we will introduce the guidelines for test point placement and quantity in such equipment.
Recently, LINPIN Instruments has successfully won the bid of Baowu Iron and Steel Company Limited's ultra-low temperature reliability environment testing project. This is another cooperation between the company and Baowu Iron & Steel following the cooperation of Anshan Iron & Steel, Hesteel, JISCO, Shagang and other iron and steel companies, which not only reflects the deep trust and tacit understanding established between the two sides in the long-term cooperation, but also lays down a colorful stroke for the diversified cooperation between the company and the iron and steel group.
Product Recommendation
Telegram WhatsApp Facebook VK LinkedIn
亚洲欧美电影一区二区_麻豆国产一区二区_免费高清在线一区_久久免费高清视频
亚洲欧美美女| 夜夜嗨av一区二区三区中文字幕| 国产精品高潮视频| 国产精品亚洲激情| 黄色成人在线网址| 亚洲精品少妇30p| 亚洲视频观看| 久久久精品国产99久久精品芒果| 欧美成人亚洲成人日韩成人| 欧美视频中文字幕在线| 亚洲男女自偷自拍| 国内精品伊人久久久久av一坑| 极品少妇一区二区三区精品视频| 亚洲欧洲日本mm| 亚洲欧美日韩直播| 麻豆久久精品| 国产精品久久久久久久久婷婷| 国产一区二区三区四区五区美女| 亚洲欧洲精品成人久久奇米网| 亚洲一级影院| 蜜桃久久精品一区二区| 国产精品乱人伦中文| 精品不卡一区| 亚洲一区影院| 亚洲一区二区成人| 久久久久久久久久久久久9999| 欧美国产精品v| 国产麻豆精品theporn| 91久久久精品| 最近中文字幕日韩精品| 亚洲欧美日韩综合aⅴ视频| 美女999久久久精品视频| 国产精品国产a| 亚洲国产欧美久久| 午夜伦理片一区| 欧美伦理影院| 国产精品地址| 亚洲国产成人高清精品| 午夜精品一区二区三区四区 | 免费中文日韩| 国产亚洲精品一区二区| 99国产精品自拍| 一本久久综合亚洲鲁鲁五月天| 久久久免费精品| 国产精品欧美日韩| 亚洲伦伦在线| 美脚丝袜一区二区三区在线观看 | 国产丝袜一区二区| 亚洲精品国精品久久99热一| 久久国产精品高清| 韩国成人福利片在线播放| 免费黄网站欧美| 一区二区高清在线| 国模精品娜娜一二三区| 欧美久久久久久久久| 欧美一区二区在线| 国产亚洲精品久久久| 亚洲一区二区三区成人在线视频精品| 国产精品黄页免费高清在线观看| 久久国产精品99久久久久久老狼| 日韩视频免费在线观看| 国产精品每日更新| 噜噜噜久久亚洲精品国产品小说| 亚洲一区自拍| 久久国产精品99国产精| 蜜桃av噜噜一区| 欧美精品免费在线观看| 国产精品久久91| 很黄很黄激情成人| 亚洲二区在线视频| 国产精品一区在线播放| 国产精品视频区| 国产主播一区二区| 欧美精品色一区二区三区| 欧美久久九九| 国产精品亚洲精品| 卡一卡二国产精品| 亚洲欧美日韩视频一区| 一区二区三区久久| 国产综合欧美| 国产精品av久久久久久麻豆网| 亚洲国产欧美日韩另类综合| 欧美日韩精品一区二区天天拍小说| 欧美成人自拍| 欧美日韩免费高清| 亚洲综合欧美| 欧美精品1区| 国产精品日日摸夜夜添夜夜av| 亚洲视频中文字幕| 欧美系列精品| 亚洲中字黄色| 国产精品免费在线| 午夜精品一区二区在线观看 | 亚洲一区国产一区| 亚洲欧洲美洲综合色网| 欧美成人一区在线| 亚洲精品一区久久久久久| 欧美精品手机在线| 一区二区日韩| 国产精品美女主播在线观看纯欲| 亚洲女人av| 国产一区二区av| 六月天综合网| 欧美日韩国产综合在线| 国模精品一区二区三区色天香| 欧美在线免费视屏| 激情亚洲网站| 欧美国产日韩视频| 在线视频一区观看| 国产精品永久入口久久久| 午夜精品一区二区三区电影天堂| 国产日本亚洲高清| 麻豆精品在线视频| 夜夜嗨av色综合久久久综合网| 国产精品国产三级国产aⅴ入口 | 国产精品一区二区三区四区五区| 久久成人在线| 亚洲日韩视频| 国产精品久久亚洲7777| 久久aⅴ国产紧身牛仔裤| 在线播放日韩| 欧美尤物一区| 国产精品久久久久婷婷| 午夜在线视频观看日韩17c| 国外成人免费视频| 欧美在线不卡| 亚洲国产精彩中文乱码av在线播放| 欧美啪啪成人vr| 午夜欧美大尺度福利影院在线看| 韩国精品久久久999| 欧美精品18videos性欧美| 亚洲欧美综合国产精品一区| 亚洲成人在线网站| 国产精品第2页| 久久这里只精品最新地址| 在线性视频日韩欧美| 黄色成人91| 欧美小视频在线观看| 卡一卡二国产精品| 亚洲欧美激情四射在线日| 亚洲国产精品久久人人爱蜜臀| 国产精品成人免费精品自在线观看| 久久久精品久久久久| 夜夜嗨av一区二区三区网页| 国产在线播放一区二区三区| 欧美日韩麻豆| 乱码第一页成人| 欧美夜福利tv在线| 亚洲免费大片| 黑人巨大精品欧美一区二区| 欧美日韩精品系列| 久久久美女艺术照精彩视频福利播放| 日韩一级黄色av| 精品白丝av| 麻豆精品国产91久久久久久| 亚洲天堂成人在线观看| 亚洲国产欧美日韩| 国产婷婷色一区二区三区四区| 欧美福利在线观看| 久久精品最新地址| 亚洲欧美精品| 亚洲毛片av在线| 依依成人综合视频| 国产区二精品视| 欧美视频免费| 欧美激情国产高清| 久久偷看各类wc女厕嘘嘘偷窃| 亚洲国产免费看| 国产亚洲精品aa午夜观看| 欧美三级视频在线观看| 猛男gaygay欧美视频| 欧美在线视频网站| 午夜欧美大片免费观看| 一本色道久久综合精品竹菊| 亚洲第一久久影院| 国产亚洲毛片在线| 国产欧美日韩综合一区在线播放| 久久野战av| 久久av一区二区三区亚洲| 亚洲淫性视频| 制服诱惑一区二区| 99精品国产热久久91蜜凸| 亚洲高清不卡av| 国模精品一区二区三区| 国产精品视频网站| 欧美午夜www高清视频| 欧美精选一区| 欧美国产极速在线| 欧美高清在线观看| 免费欧美在线视频| 免费不卡在线观看| 免费不卡亚洲欧美| 欧美xxx成人| 欧美国产日韩一区二区| 欧美freesex8一10精品| 免费看av成人| 媚黑女一区二区| 欧美1区视频| 欧美黄在线观看| 欧美96在线丨欧| 欧美成人在线网站| 欧美精品精品一区| 欧美日韩播放| 欧美日韩综合另类|